Ask Question
9 September, 06:09

A 20 foot length of two-conductor 26 AWG copper wire is used to connect a single strain gage into a quarter-bridge circuit. If each conductor has a resistance of 20.4 Ω per 1000 feet, how much has the apparent gage resistance changed? Calculate the strain induced by this error if the initial gage resistance is 120 Ω and 350 Ω.

+2
Answers (1)
  1. 9 September, 06:34
    0
    the error in conductor 1 (120 Ω) is 0.33%

    the error in conductor 2 (350 Ω) is 0.11%

    meanwhile, the current is constant.

    Explanation:

    Given:

    initial gage resistance = 120 Ω and 350 Ω

    conductor = 26 AWG

    resistance = 20.4 Ω and 1000 feet

    if 20 feet length of two conductor, each conductor has a resistance equal to

    (20.4/1000 ft) Ω

    the change in resistance in conductor is equal to

    (20 * 20) / 1000 = 400/1000

    thus the error is 0.4 Ω

    the error in conductor 1 (120 Ω) is

    (0.4/120) * 100 = 0.33%

    the error in conductor 2 (350 Ω) is

    (0.4/350) * 100 = 0.11%

    meanwhile, the current is constant.
Know the Answer?
Not Sure About the Answer?
Find an answer to your question 👍 “A 20 foot length of two-conductor 26 AWG copper wire is used to connect a single strain gage into a quarter-bridge circuit. If each ...” in 📗 Engineering if the answers seem to be not correct or there’s no answer. Try a smart search to find answers to similar questions.
Search for Other Answers